Type
Notification
Possible sources of
error
Remedy
W222
Internal voltage 24V_IO
voltage out of range
The module connected
to LD connector is
defective
• Use another module, if possible
The cable connected to
LD connector is
defective
• Use another cable, if possible
Short circuit or
overload in the 24V_IO
supply
• Contact customer service
W223
Internal voltage 24V_TMP1
voltage out of range
Short circuit or
overload in the
24V_TMP1 supply
• Contact customer service
W224
Internal voltage
24V_PRESSURE (Pirani)
voltage out of range
Short circuit or
overload in the
24V_PRESSURE
(Pirani) supply
• Contact customer service
W240
V15V out of range
Short circuit at
preamplifier
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
W241
Voltage -15V out of range
Short circuit at
preamplifier
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
E242
+15V or -15V voltage
shorted
Short circuit at
preamplifier
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
W250
REF5V voltage out of range
Short circuit at
preamplifier
• Contact customer service
VI-Board or MSB
defective
• Contact customer service
E252
REF5V voltage shorted
Short circuit at
preamplifier
• Contact customer service
VI-Board or MSB
defective
• Contact customer service