Type
Notification
Possible sources of
error
Remedy
W600
Calibration factor too low
Incorrect value entered
at calibration
• Repeat the calibration
Incorrect specimen
inserted
• Repeat the calibration
ZERO measurement
error
• Repeat the calibration
W601
Calibration factor too high
Incorrect value entered
at calibration
• Repeat the calibration
Incorrect specimen
inserted
• Repeat the calibration
ZERO measurement
error
• Repeat the calibration
W605
Signal of calibration leak too
low
Incorrect value entered
at calibration
• Repeat the calibration
Incorrect specimen
inserted
• Repeat the calibration
ZERO measurement
error
• Repeat the calibration
W630
Calibration request
Operation mode or
mass has changed
• Perform a calibration
W660
Calibration - Offset too high
Calibration leak during
zero measurement in
chamber
• Repeat the calibration
Background too high
• Use the purge function to reduce the
background
E661
Calibration - Signal too low
or offset too high
Calibration leak during
zero measurement in
chamber
• Repeat the calibration
Signal of calibration
leak too small
• Use a different calibration leak
E709
Temperature of basic unit
too low
Ambient temperature is
too low
• Increase the temperature in the
environment where the device is located
W710
Temperature of basic unit
too high
The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located
E711
Maximum temperature of
basic unit exceeded
The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located