6 | Operation
INFICON
44 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
If you repeat the measurements with the same test object, the measurement results
may differ. This is usually due to a reduced amount of solvent caused by the previous
measurement.
6.5 Purge device
► Navigate to the page "Diagnosis
→ Purge
The device automatically runs a purge process after this function is started. Here, the
test chamber and vacuum system are cyclically pumped out and ventilated so that the
background is reduced in the device following contamination.
The duration of the purge process can be adjusted.
6.6 Measurement data and device information
6.6.1 Bringing up measurement data
1
> Measurements
ð
The measurements performed are displayed in short form line by line.
2
To display the detailed view of a measurement, tap on an entry and then on the
displayed symbol
.
ð
All information stored for this measurement is displayed.
6.6.2 Transferring measurement data
Measurement results are automatically saved in the device. The last 500,000
measurements are saved. You can transfer measurement data from the internal
memory to a connected USB flash drive.
ü
Operator or Supervisor rights
1
To transfer data from the internal memory, connect a USB flash drive with
FAT32 formatting to any of the USB ports of the device.
2
> Measurements
3
Save
.
ð
All measurement data are transferred. There is an indication when the export is
completed. The measurement data remain saved on the device.
6.6.2.1 Transferring analysis data
The device records data in the internal memory for every measurement and in the
event of error.