7 | Warning and error messages
INFICON
54 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
Type
Notification
Possible sources of
error
Remedy
W125
I/O module no longer
connected
Connection to I/O
module interrupted
• Check the connection to the I/O module
• Replace the connection cable to the I/O
module
I/O module defective
• Replace the I/O module
I/O module connection
on the device defective
• Contact customer service
W127
Wrong bootloader version
The bootloader is not
compatible with
application
• Contact customer service
E129
EEPROM contains data
from wrong device class
The software of the
basic unit does not
match the EEPROM
• Contact customer service
The EEPROM does not
match this device class
• Contact customer service
W151
No communication with
operating unit
A software update or a
parameter reset has
been executed
• Confirm the warning message
• Check that the message does not appear
when you restart the device
Internal connection
problem between the
basic unit and the
operating unit
• Contact customer service
W153
Operating unit software
version is obsolete
A more up-to-date
operating unit software
is available. For
trouble-free operation,
it is recommended to
update the operating
unit software.
• Contact the customer service for the
latest operating unit software
W171
CU1000 not supported
A CU1000 cannot be
used with this device
• Disconnect the CU1000 from this device
E173
Incorrect ID in GDU
Problem in GDU
• Contact customer service
E174
GDU software is obsolete
Problem in GDU
• Contact customer service
E175
No communication with
GDU
Problem in GDU
• Contact customer service
E176
GDU not in measuring mode Problem in GDU
• Contact customer service
W190
Detector contaminated
Problem in GDU
• Contact customer service
Summary of Contents for ELT3000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 73: ......
Page 74: ......