7 | Warning and error messages
INFICON
60 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
Type
Notification
Possible sources of
error
Remedy
E399
GDU - TMP error 007 Error
in motor stage or actuation
Problem in GDU
• Contact customer service
E500
Pressure sensor p1 not
connected
Pressure sensor not
connected or cable
defective
• Contact customer service
Internal IO module
defective
• Contact customer service
E502
Pressure sensor p2 not
connected
Pressure sensor not
connected or cable
defective
• Contact customer service
Internal IO module
defective
• Contact customer service
E504
Pressure sensor p3 not
connected
Pressure sensor not
connected or cable
defective
• Contact customer service
Internal IO module
defective
• Contact customer service
W580
Maximum evacuation time
exceeded
Gross leak at test
object or the
connection to the test
object
• Check the tightness of the connection
between the leak detector and the test
object
• Use another test object, if possible
The settings value for
the max. gross leak
evacuation time is too
low
• Check and change the max. gross leak
evacuation time if necessary
W581
Maximum evacuation time
until measurement
exceeded
Gross leak at test
object or the
connection to the test
object
• Check the tightness of the connection
between the leak detector and the test
object
• Use another test object, if possible
The settings value for
the max. evacuation
time until measurement
is too low
• Check and change the max. evacuation
time until measurement if necessary