6.7.3 Updating the software of the Gas Detection Unit
Install software updates using a USB flash drive.
NOTICE
Loss of data due to disconnection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated.
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to the USB port on the device.
3
> Update > Gas detection unit
ð
At the top in the window, the active software version of the basic unit is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
ð
After completion there is an automatic restart of the system.