6 | Operation
INFICON
40 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
6.2.5 Switch on Automatic Login
You can specify whether a user of your choice is automatically logged in after the
device is switched on without the login window.
ü
Supervisor rights
ü
The requested user has already been created. See "Select, modify, create user
37]".
1
> User accounts > Manage Automatic Login
2
Enter the name of the user in the "Name" window. The input is case-sensitive.
3
Enter the current PIN of the user profile in the "PIN" window.
4
Activate the option "Active" in the window "Auto Login".
5
Save
.
6.2.6 Changing the volume
In addition to the visual display of the measurement result, a beep is sounded. You
can change the volume of the beep.
WARNING
Risk of injury due to loud noise emissions
The device can emit sounds up to a level of 100 dB (A) at the highest volume setting.
► Set the volume up to a maximum of "10".
► Use suitable hearing protection at volume settings above "5".
ü
Operator or Supervisor rights
1
> Audio
2
Adjust.
3
Save
.