4 | Description
INFICON
22 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
4.4 Factory settings
Parameter
Factory setting
Auto login
On
Default user
Supervisor
Supervisor PIN (default)
1111
Pre-set product
Universal
Measuring time
4 seconds
Measuring mass
59
Calibration mass
59
Leak threshold value
1.00E-5 mbar*l/s
Warning threshold value
8.00E-6 mbar*l/s
Automatic start of measurement
On
Volume
2
Optional input field
Off
Pre-ZERO
2 seconds
ZERO
4 seconds
Pre-LD
2 seconds
LD
2 seconds
Chamber purging time
5 seconds
Chamber vent time
4 seconds
Chamber vacuum limit
4.5 mbar
Evacuation timeout
120 seconds
Summary of Contents for ELT3000