INFICON
Warning and error messages | 7
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
53 / 74
Type
Notification
Possible sources of
error
Remedy
W106
Several EEPROM
parameters have been
initialized
A software update
introduced new
parameters
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameters corresponds to your
application
The EEPROM has
been replaced in the IO
module
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameters corresponds to your
application
The EEPROM in the
internal IO module is
defective
• Confirm the warning message
• Check if the message occurs each time
when you restart the device
• Contact customer service
E107
Internal IIC communication
error
Internal IIC
communication error
• Contact customer service
W110
Real-time clock was reset!
Enter date and time
The real-time clock has
not been set
• Enter the correct date and time
• Check that the message does not appear
when you restart the device
Battery is discharged or
defective in internal IO
module
• Contact customer service
Real-time clock
defective
• Contact customer service
W122
No response from bus
module
Connection to BUS
module interrupted
• Check the connection to the bus module
• Replace the connection cable to the bus
module
Bus module defective
• Replace the bus module
Bus module connection
on the device defective
• Contact customer service