Designation
Recommendation
Comment
Required
Optional
Proximity switch
It is possible for the
measurement to be started
immediately once the
chamber is closed by
means of a proximity
switch.
There is a M12 plug on the
back of the Gas Control
Unit.
X
Filler
No conductive material;
ceramic, glass,
polypropylene blocks
Fill up a large net volume
with filler in order to
shorten the measuring time
and increase sensitivity.
X
Ideally: Fill test chamber
with test objects to
maximum capacity.
Insulation
Butyl, ceramic, glass or
deep-drawn polypropylene
to cover the walls
Insulate the chamber walls
to prevent short circuits of
battery cells. Do not use
any adhesive!
X
Opening angle
Cover opening angle
100-110°
X
Opening aid
For heavy covers
e.g. with gas-operated
springs
X
Lock for cover
For heavy covers
Avoid risk of crushing and
cutting injuries through
design!
X
Equipment feet
Rubber feet
Use anti-slip rubber feet!