6 | Operation
INFICON
50 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
6.8.2 Calibration equipment
The following calibration equipment is available for the device:
• (Catalog Number 600-105).
The leak rate can be calibrated with the calibration equipment.
6.9 Restoring factory defaults
You can restore the device to factory settings.
Loss of settings and measurement data
After resetting to factory defaults, only the manufacturer factory settings are stored in
the memory of the device.
► Back up important settings and measurement data beforehand on a USB flash
drive. See Save user and product data and Transferring measurement data.
ü
Supervisor rights
►
> Reset device
6.10 Advanced settings
►
> Measurement
Improper changes can result in faulty measurements.
► Only make changes to the settings on this page after consultation with INFICON.
6.11 Bringing up active errors and warnings
Active errors
Errors or warnings are displayed on the active user interface. In addition, the
diagnosis symbol changes color
.
1
> Errors and warnings
ð
The "Errors and warnings" button is only available while errors or warnings
are active. Errors and warnings are displayed in list form.
2
To perform measurements, confirm active errors or warnings with the "Clear"
button.
ð
The information displayed is closed.
See also "Warning and error messages [
52]".