7 | Warning and error messages
INFICON
52 / 74
ELT3000-Battery Leak-Detector-Operating-instructions-mina95en1-10-(2007)
7 Warning and error messages
During operation, the display shows information that helps you operate the device.
Measurement values are displayed along with current device modes, operating
instructions as well as warnings and error messages. The device is equipped with
extensive self-diagnostic functions. If the electronics detect a faulty state, the device
will show this as far as possible on the display and will interrupt operation.
Warnings
Warnings warn of device states that can impair the accuracy of measurements. To
perform measurements, confirm active warnings with the "Clear” button.
Error messages
Errors are events that force the interruption of the operation. The error message
consists of a number and a descriptive text. Once you have rectified the cause of the
error, continue operation by pressing the button "Clear".
Touchscreen
You will find an overview of possible errors and warnings on the touchscreen:
►
> Help > Errors and warnings
7.1 List of warning and error messages
Type
Notification
Possible sources of
error
Remedy
W102
Timeout during
communication with
EEPROM in internal IO
module
The EEPROM in the
internal IO module is
defective or not present
• Contact customer service
W104
One EEPROM parameter
has been initialized
A new parameter was
introduced by a
software update
• Confirm the warning message
• Check that the message does not appear
when you restart the device
• Check whether the factory setting of the
new parameter corresponds to your
application
The EEPROM in the
internal IO module is
defective
• Confirm the warning message
• Check if the message occurs each time
when you restart the device
• Contact customer service
Summary of Contents for ELT3000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 73: ......
Page 74: ......