6.8 Calibrate device
6.8.1 Calibration
General calibration
Calibration is required, if:
• Operational needs require a daily calibration.
• The measuring chamber was changed.
• The ambient conditions require it.
Start calibration
ü
You have the required rights.
ü
You have an E-Check (DMC)
► In standby mode, tap the calibration icon.
ð
The calibration interface opens.
ü
The leak rate matches that of the E-Check (DMC).
ü
The vacuum chamber is not filled.
ü
The vacuum chamber is closed.
► Start an empty measurement.
ð
The empty measurement is finished
ü
The E-Check (DMC) is located in the vacuum chamber.
ü
The vacuum chamber is closed.
► Start the measurement.
ð
The second measurement, with E-Check (DMC), is completed.
At the end of the measurement of the E-Check (DMC), the new calibration factor is
determined and displayed by the device.