Operation | 6
Contura® S400-Operating-instructions-jina91en1-09-(2001)
33 / 68
6.4 Measuring
CAUTION
Hand injuries warning
Only open and close the measurement chamber when your fingers are outside the
measuring chamber halves and outside the pivoting range of the measuring chamber.
NOTICE
Damage by improper filling of the measuring chamber
Leaking liquids entering the tubing, can interfere with the function of the device. Sharp,
fats or oils may damage the textile mesh, membrane, spacer diaphragm ring and
sealing lips.
► Only check dry packaging or packages that do not contain liquids.
► Avoid contamination of the measuring chamber by oils, fats, or hydrocarbons.
► Do not use sharp edged objects without protective frame in the measuring chamber.
Avoid measuring inaccuracies:
► Place packages so that the gaskets of the measuring chamber halves are not
covered or touching!
► Avoid measuring test objects with significant differences in temperature to the
surroundings!
► Keep the gaskets of the measuring chamber halves clean. If you do not remove
contaminants, measuring results may be distorted.
ü
You have made general settings, see Basic settings [
ü
You have saved the settings for the desired product in the device, see Product-
24].
ü
You have selected the desired product, see Select product [
31].
ü
You have carried out a ZERO measurement before changing a product line and at
regular intervals, see "Perform ZERO measurement [
31]”.
1
Call the measurement screen.
ð
The measurement screen appears automatically after a user logs in.
Alternatively, press
.
2
Place the test object in the measuring chamber.
3
Close the measuring chamber and start the measurement. Regarding the start
options, also see the descriptions in Automatic measurement starting ON or