6.3 Settings for the measurements
6.3.1 Select product
1
> Products
ð
Existing products are displayed. If you cannot fine the desired product, you
can create it, see Select, modify or create product (measurement settings)
2
Click the desired product name and select.
3
Load
.
ð
You will be asked to perform a ZERO measurement.
6.3.2 Perform ZERO measurement
To compensate for external factors such as air pressure or temperature on the
measurement, perform ZERO measurements in the following cases as described
below:
► After a prompt on the display
► Before starting a series of measurements
► At least once an hour in measuring mode
1
Select
>
.
2
Empty the measuring chamber.
3
Start the ZERO measurement.
ð
The result is displayed in green and the values are taken over. In the event of a
fault, the result is displayed in red.
6.3.3 Using the input field in the measurement window
If required, you can set up an input field in the measurement window to enter an
additional information text there. For example, a batch number.
This text is not only displayed in the measurement window, but is also logged during
data recording. The text is assigned to the measurement performed.
1. Setting up the input field for the measurement window
1
> Device
2
Activate the "Optional input field".
3
Save
.