Compare the data displayed there for the measured test specimens with the
known dimensions of these test specimens.
Test specimen bar red: Outer volume 100 ml
Test specimen bar blue: Outer volume 50 ml
Test specimen bar green: Outer volume 50 ml
A deviation of approx. 5 ml is within the range of the tolerance and does not affect the
gross leak detection.
6.11 Call active errors and warnings
Active errors
Errors or warnings are displayed on the active working surface. In addition, the
diagnostics symbol changes color
.
1
> Errors and warnings
ð
The "Errors and Warnings" button is only available while errors or warnings
are active. Errors and Warnings are displayed in list form.
2
To perform measurements, confirm active errors or warnings with the "Clear"
button.
ð
The information displayed is closed.
See also "Warning and error messages [
6.12 Logging off from the device
1
Press on your name which appears on the top right of the display.
ð
The "User options" window opens.
2
You log off from the device via the button "Log off".
ð
The login window opens.
6.13 Switching off the device.
You can switch off the device at the mains switch at any time. The parameters set in
the device remain saved.