6 | Operation
INFICON
38 / 68
Contura® S400-Operating-instructions-jina91en1-09-(2001)
NOTICE
Data loss due to an aborted connection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to the USB port on the device.
3
> Update > Update Basic Unit
ð
At the top in the window the active software version of the basic unit is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
ð
After completion there is an automatic restart of the system.
6.6.3 Update the software in expert mode
1
> Update > Update Operating Unit > Experte
ð
Software versions already available on the device are shown in a list.
2
Highlight any of the software versions as required and proceed to step 5.
ð
Resetting to an older software version is possible.
3
If alternatively you want to add a new software version, connect a FAT32
formatted USB flash drive with the update file to one of the USB ports of the
device.
4
In order to load the new software version, press on
.
5
To activate the new version, select the desired list entry and press
.
ð
After completion there is an automatic restart of the system.
6.7 Save user and product data
You can save all user and product data to a USB flash drive and restore whenever
necessary.
No backup of measurement data
When backing up user and product data no measurement data backup is performed.
To backup measurement data see "Transferring measurement data [
35]”.
ü
Supervisor - rights
1
Insert a FAT32-formatted USB flash drive into a USB port on the device.