INFICON
Operation | 6
Contura® S400-Operating-instructions-jina91en1-09-(2001)
37 / 68
ð
All measurement data is deleted.
6.5.6 Call device information
►
> Device information
ð
The stored information will be displayed.
6.5.7 Call log
Button to display device messages in list form. This information is useful when you
contact the manufacturer's service.
►
> Log
6.6 Updating the software
The device has two different software versions: One for the user interface and one for
the basic unit. Each has its own independent versions number.
6.6.1 Update the user interface software
Import software updates using a USB flash drive.
NOTICE
Data loss due to an aborted connection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
1
Copy the file into the main directory of a FAT32 formatted USB flash drive.
2
Connect the USB flash drive to a USB port on the device.
3
> Update > Update Operating Unit
ð
At the top in the window the active software version of the user interface is
shown.
If one or more versions of the software are on the USB flash drive the most
recent version is shown on the line below. If this is the same as the version
already installed the background is green, otherwise it is red.
4
In order to load the new software version, press on the button "Update".
ð
After completion there is an automatic restart of the operating unit.
6.6.2 Updating the software of the basic unit
Import software updates using a USB flash drive.