4 Installation
4.1 Setup
• In order not to distort the measurement results, select a location where the
possible room temperature for the device is constant.
• In order not to block the exhaust opening on the lower part of the device, place the
feet of the unit on a firm even surface.
• To easily reach the power switch on the back of the device, ensure there is
sufficient free space behind the device.
• Do not expose the device to direct sunlight.
WARNING
Danger from moisture and electricity
Moisture entering the device can lead to personal injury due to electric shocks as well
as damage to property due to short circuiting.
► Only operate the device in a dry environment.
► Operate the device away from sources of liquid and moisture.
CAUTION
Risk of injury from lifting the heavy device
The device weighs approx. 54 kg and can slip out of your hands.
► Lift and transport the device only with two people.
► To lift get hold below the floor panel.
NOTICE
Material damage from overheated device
The device heats up during operation and can overheat without sufficient ventilation.
► Please note the technical specifications.
► Ensure sufficient ventilation, especially on the ventilation slots on the left and right
of the device: There should be free space in the front, to the rear and sides of the
unit of at least 10 cm.
► Keep heat sources away from the device.