INFICON
Operation | 6
Contura® S400-Operating-instructions-jina91en1-09-(2001)
27 / 68
6.2.6.2 Select, modify or create product (measurement settings)
For setting measurement specifications you need a product that has been created.
ü
Operator or Supervisor rights
1
> Products
ð
Existing products are displayed as a list.
2
Press a product name and choose from the displayed tool-bar or from the
display:
, to load a product.
, to modify product settings.
The window "Product Settings" will open.
, to copy product settings.
The "Product Settings" window opens with the entries of the copy source.
, to delete a product.
A confirmation screen appears.
, to create a new product.
The window "Product Settings" opens.
3
After using some tools, the window "Product Settings" opens. In this case, enter
under "Product Name" a product name as required, change it or keep it.
Product name
Measuring mode
Supervisor
Product settings
Part 1
Part 2
Chips
Universal
mbar l/s
Max. leak rate
4
Select the "Measuring Mode", see "Determine the Measuring mode [
25]”.
5
Under "Max. Leak rate", enter the value at which the product is to be reported as
"leaking". The default setting is 0.01 mbar l/s.