6 | Operation
INFICON
24 / 68
Contura® S400-Operating-instructions-jina91en1-09-(2001)
6.2.4 Switch off Automatic Login
Factory setting
As per factory settings, after switching on the device the user "Supervisor"
automatically logs in and the measurement screen is called. This default user also has
the permissions of the group "Supervisor". Without changing this setting, any user can
operate all functions without restriction.
You can specify that the login window appears after you turn on the device instead of
automatically logging on.
In the login window, all users can login, which have been already registered in the
device, see"Select, modify, create user profile [
ü
Supervisor - rights
1
> User accounts > Manage Automatic Login
2
Deactivate in the window "Auto Login" the option "Active".
3
Save
.
ð
After restarting the device, the current settings are applied.
6.2.5 Switch on Automatic Login
You can specify if a user of your choice is automatically logged in after the device is
switched on without the login window.
ü
Supervisor - rights
ü
The requested user was already created. See "Select, modify, create user profile
1
> User accounts > Manage Automatic Login
2
Enter the name of the user in the "Name" window. Note the uppercase/
lowercase.
3
Enter the current PIN of the user profile in the "PIN" window.
4
Activate in the window "Auto Login" the option "Active".
5
Save
.
6.2.6 Product-specific settings
In the next two chapters the setting of the "measuring mode" and details on the
product such as "Max. leak rate” will be handled. If you still have questions left after
reading, please contact us.