hydratight
76
Note:
In order for the BOLTSCOPE PRO to display the correct
measurement value during calibration, the material velocity and
temperature coefficient must be set to the same values as in the calibration
section.
32) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu
items until
VELOCITY
is highlighted.
33) Press the
ENTER
key to display the
Digits Edit Box
.
34) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
35) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
36) Repeat steps
34 & 35
until the
VELOCITY
value is correctly displayed.
37) Press the
OK
key to return to the
MATL
menu, or
ESC
to cancel
entering the
VELOCITY
.
38) Repeat steps
32 – 37
to enter the
TEMP COEF
.
39) Repeat steps
32 – 37
to enter the
STRESS FACTOR to 1.000”
.
40) Proceed to the next section.
Before the probe zero is calculated the cal bars / glass block needs to be
ultrasonically measured. Once the zero value has been calculated, the results
will then need to be documented. This procedure is demonstrated in the next
section.