the bolt, thus causing potentially erroneous measurements. Be consistent
and as methodical in your methods as possible. This will help to avoid
transducer placement errors.
1) Press the
UP
,
DOWN
,
LEFT
, and
RIGHT
arrow keys to scroll the target
cell cursor to the desired storage location.
Note:
Elongation values must be stored in column B - ZZ.
Note:
Do Not press the
AUTO SET
key, while measuring elongation’s, as
this activates a high speed mode used specifically with our shut-off box.
The
AUTO SET
feature is only used when measuring reference lengths. If
the
AUTO SET
key is accidentally pressed, press it again to turn the high
speed mode off.
2) Press the
ENTER
key to save the current elongation in the highlighted
cell location.
Note:
If the measurement was saved in error, press the
CLR
key at
anytime to clear the stored reading and re-measure using steps
1 & 2
above.
3) Continue to repeat this process until all the elongation measurements
have been completed.
Note:
The
Group View Box
can be aborted at any time by pressing the
MEAS
key.