4) Use the
UP
and
DOWN
arrow keys to scroll through the groups until the
target group is highlighted.
5) Press the
ENTER
key to activate the confirmation screen.
6) Press the
OK
key to load the group from memory.
7) Press the
MEAS
key to return to the measure screen.
8) Press the
MEAS
key once to activate measure menu items. Press the
MEAS
key multiple times to move right and the
ESC
key multiple times
to move left until the
LOC
cell is highlighted.
9) Press the
ENTER
key to display the
Group View Box
.
10) Press the
UP
,
DOWN
,
LEFT
, and
RIGHT
arrow keys to scroll the
stored readings and corresponding display view. Notice as the cursor is
moved to a different cell, the display will be updated with the display
view saved with the reading.
11) Scroll the cursor to the next blank cell for the probe zero bolt. This will
always be a cell located in columns
B – ZZ
. As soon as the user has
scrolled to blank cell, the
BOLTSCOPE PRO
will switch into elongation
mode and use all the data from the probe zero bolt initially stored.
12) Press the
MEAS
key to return to the measure screen.