BOLTSCOPE PRO
- Bolt Tension Monitor
25
Threshold:
Enables the user to set the sensitivity level of the
BOLTSCOPE
PRO
. The amplitude of the signal must reach and exceed the threshold level
before a measurement is detected. Refer to section 12.5 for further info.
Polarity:
The
BOLTSCOPE PRO
operates on a zero crossing detection
principal. This feature toggles which stroke of the cycle the crossing detection
uses, either positive or negative. Refer to 13.3 for further info.
Pulse:
The
BOLTSCOPE PRO
has an adjustable pulse width for both high
penetration and resolution requirements. The pulse width refers to the duration
of time the pulser is on. The options are
Spike
,
Thin
, and
Wide
. Refer to 13.1
for a further explanation.
Rect Waveform:
This option provides the user an outlined or filled view option
when the display setting is in RECT (rectified) wave mode only. Refer to 13.4 for
further info.
4.7 AUTO – Menu
Auto Set:
Automatically locates the detection point if the measurement is out of
the viewable display area. Used when measuring reference lengths only. Refer
to section 12.9 for further info.
Approx Len:
In order for the user to use the Auto Set feature, an approximate
length of the fastener must be entered. The approximate length gives the
BOLTSCOPE PRO
some idea of where to turn on the receiver and look for the
detection, or end of the bolt. This sub menu item enables the user to edit the
effective length at anytime. Refer to section 12.9 for further info.
Measure Mode:
Used to select the measurement mode for different application
requirements. The mode options are P-E (pulse-echo), or P-E GT (pulse-echo
w/gate).
Gate:
Gates allow the user to view areas or sections of the waveform and ignore
others. Refer to section 12.6 for further info.