BOLTSCOPE PRO
- Bolt Tension Monitor
85
counter clockwise applying a small amount of pressure to eliminate any
excess couplant between the transducer and cal standard.
Be sure to
always place the transducer in the same location
.
This will help to
eliminate any potential measurement errors caused by changing the
sound path.
7) Press the
MEAS
key once to activate the measure menu items. Press
the
MEAS
key multiple times to move right and the
ESC
key multiple
times to move left, until the
LOC
cell is highlighted.
8) Press the
ENTER
key to display the
Group View Box
.
9) Press the
UP
,
DOWN
,
LEFT
, and
RIGHT
arrow keys to scroll the target
cell cursor to the desired storage location.
Note:
Since this is the 1
st
verification following the initial measurement of
the
Probe Zero Cal Bars
, the cursor should be located on a cal standard in
column B
. It’s also important to note that the
L-REF
stored in
column A
,
also contains all the
BOLTSCOPE PRO
settings when the
L-REF
was
initially measured.
These settings
will automatically be copied to all the
columns B-ZZ, for each individual bolt
. Therefore,
no adjustments
to
the
BOLTSCOPE PRO
should be necessary.
10) If the elongation is +/- .0001” and “In Spec”, Press the
ENTER
key to
save the current reading in the highlighted cell location, storing all the
current probe zero verification data.
Repeat steps 6 - 19 for the other
probe zero calibration standards. If the probe zero is “Out of
Spec”, proceed to the next step.