BOLTSCOPE PRO
- Bolt Tension Monitor
77
Setting up the BOLTSCOPE PRO for Measurement
Note:
Check to see if the system was previously calibrated. You should
NOT
change the calibration unless the transducer, transducer cable, or
instrument has been changed. In any event, the probe zero calibration
should always be checked prior to changing it and only changed if it is
required, as you can introduce error by uncertainty during the zero
procedure. If the
BOLTSCOPE PRO
has been previously calibrated, open
the calibration data saved in the
BOLTSCOPE PRO
and check the probe
zero by simply measuring the bars in an empty location (columns B – ZZ) in
the existing probe zero data group.
If the zero is slightly off, minor
adjustments to the temperature, located in the Main Measure Menu,
should be all that’s necessary to bring the error between the two bars within
a tolerance of .0005”
.
1) With the menu items still active from the previous section, press the
MENU
key multiple times to tab right and the
ESC
key multiple times to
tab left until the
GEOM
menu is highlighted and displaying the submenu
items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
QUANTITY
is highlighted.
3) Use the
LEFT
and
RIGHT
arrow keys to scroll the quantity options until
the
ELONG
option is displayed.