hydratight
82
Note:
Lower temperatures will make the bolt read longer, and higher
temperatures shorter.
28) Repeat steps 14 – 25 above to re-calculate the zero value using the
adjusted temperature setting.
Note: The above steps may need to be repeated a couple of times in order
to bring the probe zero calibration into specification - .0005”. Note: This is
typically not necessary when using the glass block as a cal standard.
29) Once the zero/calibration is within specification, these initial lengths
should be saved in column A of the zero group previously created for
future reference and verification.
Note:
A periodic check should be conducted to prior to starting a project, to
confirm the zero is still accurate. If it’s determined the
BOLTSCOPE PRO
is out of spec, only minor adjustments to the temperature should be
necessary. Refer to the section on
“Verifying the probe zero calibration”.
Note:
Now that the zero is within spec, the values should be documented
in the probe zero calibration group created and opened in an earlier section.
30) Press the
MEAS
key twice to activate measurement screen. Press the
MEAS
key multiple times to move right, and the
ESC
key multiple times
to move left until the
LOC
cell is highlighted.
31) Press the
ENTER
key to display the
Group View Box
.
32) Press the
UP
,
DOWN
,
LEFT
, and
RIGHT
arrow keys to scroll the target
cell cursor to the desired storage location.