Interfaces and data communication
Chapter 6
OPERATION MANUAL
V1.2.0en/11.12.20//0.3.1
KISS Immersion Circulator
45
6.3.1.3
Command “G” (General)
This command transmits the most important temperatures and status information in a cycle. A
modified setpoint is not stored in the permanent memory, i.e. this value is lost when switching off
the machine.
Byte
ASCII
Hex
Description
Master sends:
[M01G0Dsattttpp\r
1. Byte
[
5Bh
Start character
2. Byte
M
4Dh
Master ID
3. Byte
0
30h
Slave address
4. Byte
1
31h
Slave address
5. Byte
G
47h
Command qualifier
6. Byte
0
30h
Length of the command: 0Dh = 13 bytes (number of bytes without
checksum and end character)
7. Byte
D
44h
8. Byte
s:
C / I / O / *
43h / 49h /
4Fh / 2Ah
Temperature control mode
Meaning of the characters in the send string:
“C” (43h) = Circulation, switch circulation on;
“I” (49h) = Turn internal temperature control on;
“O” (4Fh) = Off, turn temperature control off;
“*” (2 Ah) = Do not change the current state.
9. Byte
a:
0 / 1 / *
30h / 31h /
2Ah
Alarm acknowledgment
Meaning of the characters in the send string:
“0” (30h) = No alarm acknowledgment;
“1” (31h) = Any pending alarm tone is acknowledged;
“*” (2 Ah) = Do not change the current state.
10. Byte t
tttt /
****
Query or set the setpoint
Meaning of the characters in the send string:
Setpoint with 16-bit resolution (2 bytes, thus 4 ASCII characters)
“tttt” =
0000h (0.00 °C) to 7FFFh (327.67 °C)
FFFFh (-
0.01 °C) to 8000h (
-
327.68 °C)
0190h corresp
onds to +4 °C, (30h, 31h, 39h, 30h)
FE70h corresponds to -
4 °C (46h, 45h, 37h, 30h)
“****” (2Ah, 2Ah, 2Ah, 2Ah) = no change to the setpoint, setpoint is
only queried
11. Byte t
12. Byte t
13. Byte t
14. Byte p
Checksum
Checksum
It is generated from bytes 1 to 13.
15. Byte p
Checksum
16. Byte \r
0Dh
End character CR
The slave responds:
[S01G15sattttiiiieeeepp\r
1. Byte
[
5Bh
Start character
2. Byte
C
53h
Slave ID
3. Byte
0
30h
Slave address
4. Byte
1
31h
Slave address
5. Byte
G
47h
Command qualifier
6. Byte
1
31h
Length of response: 15h = 21 Bytes
7. Byte
5
35h
8. Byte
s:
C / I / O
43h / 49h /
4Fh
Temperature control mode
Meaning of the characters in the response string:
“C” (43h) = Circulation, circulation is on;
“I” (49h) = Internal temperature control is on;
“O” (4Fh) = Off, temperature control is off.
Structure Command
“G” (General)