Service/maintenance
Chapter 7
OPERATION MANUAL
V1.2.0en/11.12.20//0.3.1
KISS Immersion Circulator
49
Code
Cause
Effect, measure
033
Error EP0 (Flash)
Please contact Customer Support.
034
Error EP1 (EEPROM)
035
Error EP2 (NVRAM)
036
Synchronization
037
Parameters not equal
038
Invalid status
039
Error safety chip
042
Pump protection activated
The pump motor is overheated.
Check the ambient conditions.
Check the viscosity of the thermal fluid.
Turn the temperature control unit off and let it cool down.
7.2
Electrical fuse
The thermal overcurrent circuit breakers for all-pole disconnection (L and N) are located at the back.
In case of a fault (no function and/or no display) please first check if the overcurrent circuit breakers
have tripped. If the overcurrent circuit breakers trigger again immediately after reversing, please
unplug the power cord and contact Customer Support immediately (the phone number can be
found on page 57 in section
).
7.3
Maintenance
Cleaning/maintenance while the temperature control unit is operating
MORTAL DANGER FROM ELECTRIC SHOCK
Stop an ongoing temperature control process.
Turn off the temperature control unit.
Also disconnect the temperature control unit from the power supply.
Performing maintenance work not described in these operation manual
MATERIAL DAMAGE ON THE TEMPERATURE CONTROL UNIT
Please contact Huber for maintenance work that is not described in these operation manual.
Maintenance work not described in these operation manual is reserved for qualified specialists
trained by Huber.
Safety-relevant components may only be replaced by equivalent ones. The specified safety
values for the respective component must be observed.
7.3.1
Function check and visual inspection
Cooling*
Description
Maintenance
interval
Comment
Person re-
sponsible
A/W
Visually inspect
hoses and hose
connections
Prior to switching
on the tempera-
ture control unit
Exchange leaking hoses and hose
connections prior to switching on
the temperature control unit.
Please see on page 50 the section
»Replacing temperature control
Operating
company and /
or operators
A/W
Check the fill level
in the collecting
container at the
>Overflow<
[12]
(if
present)
Prior to switching
on the tempera-
ture control unit
Check the fill level in the collecting
container. Empty as required.
Follow the instructions for the
disposal of thermal fluid on page
Operating
company and /
or operators
Control intervals