Interfaces and data communication
Chapter 6
OPERATION MANUAL
V1.2.0en/11.12.20//0.3.1
KISS Immersion Circulator
47
Function
Master sends
Slave responds
Description
Read the set-
point
SP?\r\n
SP +02500\r\n
The setpoint is set to 25.00 °C.
Read the inter-
nal actual value
TI?\r\n
TI +02499\r\n
Currently, the internal actual value is 24.99 °C.
Read the exter-
nal actual value
TE?\r\n
TE +02499\r\n
Currently, the external actual value is 24.99 °C.
TE -15100\r\n
An external sensor is not connected or does not
exist.
Read the tem-
perature control
mode
CA?\r\n
CA +00000\r\n
Temperature control and circulation are inactive.
CA +00001\r\n
Temperature control and circulation are active.
Function
Master sends
Slave responds
Description
Setting the
setpoint
SP@ -01234\r\n
SP -01234\r\n
The setpoint is set to -
12.34 °C.
Starting the
temperature
control unit
CA@ 00001\r\n
CA +00001\r\n
The temperature control process is started.
Stopping the
temperature
control unit
CA@ 00000\r\n
CA +00000\r\n
The temperature control process is stopped.
Available read com-
mands
Available write com-
mands