Commissioning
OPERATION MANUAL
Chapter 2
KISS Immersion Circulator
V1.2.0en/11.12.20//0.3.1
26
2.9.3
Installing collecting container
PROCEDURE
Install a suitable hose at the
>overflow<
[12] on the temperature control unit (if present). The
hose must be compatible with the thermal fluid and the temperature.
Put the other end of the hose in a suitable collecting container.
2.9.4
Connecting the functional earth
PROCEDURE
If required, connect the temperature control unit’s
>Functional earth terminal<
[87] to the
building’s grounding point. Use a ground strap for this purpose. For the exact position and thread
size please refer to the wiring diagram from page 58 in Section
.
2.10
Connecting externally closed application
The illustration “connection diagram” can be found on page 58 in section
.
2.10.1
Connecting an externally closed application
Overpressure in the application (e.g. > 0.5 bar (g) with glass apparatus)
MATERIAL DAMAGE TO THE APPLICATION
Provide an overpressure protective device to prevent damage to the application.
Do not install valves/quick-release couplings in the feed/discharge lines from the temperature
control unit to the application and from the application to the temperature control unit.
If valves/quick-release couplings are required:
Install burst disks on the application itself (at the feed and discharge lines).
Install a bypass upstream of the valves/quick-release couplings for the application.
Matching accessories (e.g. bypasses to reduce pressure) can be found in the Huber catalog.
To enable your application to be operated correctly and eliminate air bubbles from the system, you
must ensure that the
>Circulation flow<
[1] connection from the temperature control unit is at-
tached to the lower connection point of the application and the
>Circulation return<
[2] into the
temperature control unit is attached to the higher connection point of the application.
PROCEDURE
Remove the screw plugs from the
>Circulation flow<
[1] and
>Circulation return<
[2] connections.
Then connect your application to the temperature control unit using suitable thermal fluid hoses.
The corresponding wrench sizes can be found in the table on page 23 in section
.
Check the connections for leaks.
2.11
Connecting to the power supply
Based on local circumstances, it may be that you need to use an alternative power cable instead of
the supplied original power cable. Do not use a power cable that is longer than
3 m
to be able to
disconnect the temperature control unit at any time from the mains. Have the mains cable only
replaced by a qualified electrician.
Example: Connecting
an externally closed
application