Commissioning
OPERATION MANUAL
Chapter 2
KISS Immersion Circulator
V1.2.0en/11.12.20//0.3.1
22
2.3.1
EMC-specific notes
Connecting cables in general
Prerequisites for a failure-free operation of the temperature control units incl. their connections with external
applications: Installation and wiring must be carried out professionally. Topics affected: “Electrical safety“ and
“EMC-compliant wiring“.
Cable lengths
For flexible/fixed cable routing longer than 3 meters, the following must amongst other things be observed:
- Equipotential bonding, grounding (see also technical data sheet “Electromagnetic compatibility EMC“)
- Compliance with “external“ and/or “internal“ lightning/overvoltage protection.
- Constructional protective measures, professional cable selection (UV resistance, steel pipe protection, etc.)
Attention:
The responsible body is responsible for compliance with national/international directives and laws. This also
includes the testing of the installation/wiring required by law or standards.
These devices are suitable for the operation in
“industrial electromagnetic environments”
. It meets the
“im-
munity requirements”
of the currently applicable
EN61326-1
, which are required for this environment.
It also meets the
“interference emission requirements”
for this environment. It is a
Group 1
and
Class A
unit
according to the currently applicable
EN55011
.
Group 1
specifies that high frequency (HF) is only used for the function of a device.
Class A
specifies the interfer-
ence emission limits to be observed.
2.4
Installation conditions
Temperature control unit is connected to the power supply line
DEATH FROM ELECTRICAL SHOCK BY DAMAGE TO THE POWER CABLE.
Do not put temperature control unit on power cable.
Operating the temperature control unit fitted with castors without brakes activated
CRUSHING OF LIMBS
Activate brakes on the wheels.
▪
Allow the temperature control unit to acclimate for about 2 hours when changing from a cold to a
warm environment (or vice versa). Do not turn on the temperature control unit before!
▪
Install upright, stable and without tilt.
▪
Use a non-combustible, sealed subsurface.
▪
Keep environment clean: Prevent slip and trip hazards.
▪
Wheels must be locked after the installation, if installed!
▪
Spilled/leaked thermofluid must be disposed of immediately and properly. Follow the instructions
on page 15, section
»Proper disposal of resources and consumables«
for the disposal of ther-
mofluid and material.
▪
Observe the floor load bearing capacity for large units.
▪
Observe the ambient conditions.
2.5
Recommended temperature control and cooling water hoses
Use of unsuitable/defective hoses and/or hose connections
INJURIES
Thermal fluid
Use appropriate hoses and/or hose connections.
Check periodically for leaks and the quality of the hose and hose connections and take suitable
measures (replace) as required.
Isolate and protect temperature control hoses against contact/mechanical load.
Cooling water
Reinforced hoses must be used to satisfy tougher safety requirements.
Shut off the cooling water supply to the temperature control unit even for shorter downtimes
(e.g. overnight).