Setup mode
OPERATION MANUAL
Chapter 4
KISS Immersion Circulator
V1.2.0en/11.12.20//0.3.1
36
thermal fluid just filled is higher than the cut-out value set for the overtemperature protection. Set
the overtemperature protection to the thermal fluid you are using. Please note: The printed scale
can deviate by - 25 K from the set cut-out value.
4.1.3.2
Setting the overtemperature protection
You need a screwdriver (flat blade 1.0 x 5.5) to set the cut-out value of the overtemperature protection.
PROCEDURE
Use a screwdriver to set the cut-off value on the potentiometer. The cut-out value must be set to
match the thermal fluid you are using. It is not required to switch on the temperature control unit.
4.1.4
Testing overtemperature protection for functionality
Overtemperature protection (OT) does not trip
MORTAL DANGER FROM FIRE
Test the response of the device every month and after each change of the thermal fluid in
order to assure proper functioning.
The steps below are carried out without permanent monitoring of the temperature control unit
DAMAGE TO AND IN THE VICINITY OF THE TEMPERATURE CONTROL UNIT
The following actions may only be carried out while constantly monitoring the temperature
control unit and the application!
The overtemperature protection is installed only in temperature control units that have a heater. You
need a sufficiently large-sized screwdriver to check the overtemperature protection for functionality.
Steps to test the correct functioning of the overtemperature protection:
PROCEDURE
Note down the cut-out value of the overtemperature protection set on the potentiometer.
Switch on the temperature control unit.
Enter a setpoint (room temperature). Further information can be found on
page 34 in section
.
Press the
>Start/Stop button<
[E]to start temperature control.
Use a screwdriver to set the new cut-off value on the potentiometer. This shutdown value must
be
below
the displayed internal temperature. The overtemperature protection is triggered.
Turn off the temperature control unit.
Use a screwdriver to reset the cut-off value on the potentiometer to the original value.
Immediately take the temperature control unit out of operation if the overtemperature protection
is not triggered. Immediately contact Customer Support. The telephone number can be found on
page 57, section
. Do not put the temperature control unit back into operation.
Setting the
cut-out value