Function description
Chapter 3
OPERATION MANUAL
V1.2.0en/11.12.20//0.3.1
KISS Immersion Circulator
33
3.5
Menu function
Pressing both
>Arrow keys<
[B] simultaneously calls up the main menu. Some menu items cannot be
selected depending on the configuration of the temperature control unit.
Display
Description
KISS
OLÉ
Setpoint 1
Sets the setpoint.
Use the
>Arrow keys<
[B]
to change the setpoint.
X
X
Adjusting
brightness
Adjusting the brightness of the OLED display.
Use the
>Arrow keys<
[B]
to change the brightness.
X
X
Sensor
configura-
tion
This menu item makes available:.
1) Adjustment of the internal sensor (input options: Offset (K))
2) Adjustment of the external sensor (input options: Offset (K))
3) Temperature unit (choose between “Celsius” and “Fahrenheit”)
4) Mode (choose between “Internal temperature control”, “Venting” and “Circu-
lation”
5) Display of external Pt100 sensor – (activating the display of an external Pt100
process display sensor)
6) Flow sensor / pressure sensor display – (activating the display of the optional
flow sensor / pressure sensor)
X
O
X
X
O
–
X
O
X
X
O
M
Interfaces
This menu item makes available:
1) RS232 1 (setting of “Baud rate” and “Mode” (HuberBus))
2) RS232 2 (setting of “Baud rate” and “Mode” (HuberBus))
3) USB device (setting of “Baud rate” and “Mode” (HuberBus))
Only the Huber service engineer is allowed to use the "STBus" mode.
4) Floating contact (selection of “Off”, “Alarm” and “Unipump/PCS”)
5. External control signal (selection between “Off”, “Setpoint2” and “Standby”)
X
X
X
–
–
X
O
X
O
O
Protection
Options
This menu item makes available:.
1) Setpoint2 (to input the second setpoint)
2) Setpoint minimum (to input the lower limit of the adjustable setpoint)
3) Setpoint maximum (to input the upper limit of the adjustable setpoint)
4) Power failure automatic (select between “Off” and “Automatic”)
–
X
X
X
O
X
X
X
System
This menu item makes available:
1) Heating output (only with temperature control units; setting in %
2) Select Language (choose between “English” and “German”)
3) Cooling bath (select between “Without cooling bath” (Off), “With cooling bath
and common power supply” (On) and “With cooling bath and separate power
supply” (On))
4) System Information (display different serial numbers (Serial Number) and
version statuses)
5) Service menu (only for Huber service technicians. This submenu is password
protected)
6) Factory settings (choose between “Continue” and “Cancel”)
X
X
M
X
X
X
M
X
–
X
X
X
X = standard, O = optional, M = model-dependent– = not possible
Main menu
Overview of the menu
items
Summary of Contents for KISS 1 Series
Page 1: ...KISS Immersion Circulator...
Page 2: ......
Page 3: ...OPERATION MANUAL KISS Immersion Circulator...
Page 4: ......
Page 10: ...OPERATION MANUAL KISS Immersion Circulator V1 2 0en 11 12 20 0 3 1 10...