1.4.2
Safety devices to DIN 12876
The rating of your temperature control unit is stated on the data sheet in the appendix.
Classification
Temperature
control medium
Technical requirements
Identification
d)
I
Non-combustible
a)
Overheat protection
c)
NFL
II
Combustible
b)
Adjustable overheat protection
FL
III
Combustible
b)
Adjustable overtemperature protection
and additional low-level protection
FL
a)
Usually water; other fluids only if non-combustible even within the temperature range of an individual fault.
b)
The temperature control media must have a fire point of ≥ 65 °C.
c)
The overheat protection can, for instance, can be realized using a suitable fill level sensor or a suitable temper-
ature limiter.
d)
Optional at the choice of the manufacturer.
Mechanical overtemperature protection
Only temperature control units with a heater are fitted with a mechanical overtemperature protec-
tion. Set the overtemperature protection as described on page 35 in section
.
Low level protection
A mechanical float is used for level monitoring. In the bath vessel, a floating body, which is guided in
a device, floats on the surface of the thermal fluid. Depending on the level of the thermal fluid, the
float device signals the electronics a
state of good
(in case of sufficient filling) or a
state of bad
(in
case of insufficient filling). The functionality of the float switch is checked at regular intervals during
continuous operation.
Rating of laboratory
thermostats and
laboratory baths
Overview of the tem-
perature thresholds