Shutting down
Chapter 8
OPERATION MANUAL
V1.2.0en/11.12.20//0.3.1
KISS Immersion Circulator
55
8.3
Draining the temperature control unit
PROCEDURE
Drain the temperature control unit as described from page 37 in section
.
8.4
Draining the cooling water
This section must be observed when using water-cooled temperature control units.
8.4.1
Draining process
Pressurized cooling water connections
RISK OF INJURY
Wear your personnel protective equipment (e.g. safety goggles).
Carefully open the cooling water connection. Open slowly (1-2 signal edges) and drain the
cooling water slowly.
Building side isolating valves are not closed
DAMAGE BY ROOM FLOODING
Close the building’s isolating valves in the cooling water supply and return lines.
PROCEDURE
Close the building’s isolating valve in the cooling water supply and return lines.
Place the collecting container below the input and output of the
>Cooling coil<
[29].
Unscrew the connection at the
>Cooling coil<
[29]. The cooling water will begin to drain from the
lines.
Remove the cooling water from the
>Cooling coil<
[29]. Allow the cooling water to fully drain to
prevent the risk of freezing during transport and storage!
8.5
Deinstalling the collecting container
PROCEDURE
Remove the hose from the collecting container.
Follow the instructions for the disposal of thermal fluid on page 15 in Section
.
Dismantle the hose at the
>Overflow<
[12].
8.6
Uninstalling an external application
PROCEDURE
Disconnect the external application from the temperature control unit.
8.7
Packing
Use the original packaging wherever possible! Further information can be found on page 20 in sec-