FULLTEST3
EN - 89
INSULATION RESISTANCE (M
:
)
DC Test voltage
[V]
Range
[M
:
]
Resolution
[M
:
]
Accuracy
Overload
protection
100
0.00
y
9.99
0.01
± (3%rdg + 3dgt)
CAT III 300 V
10.0
y
20.0
0.1
20.0
y
99.9
± 5%rdg
250
0.00
y
9.99
0.01
± (3%rdg + 3dgt)
10.0
y
20.0
0.1
20.0
y
99.9
± 5%rdg
100
y
250
1
500
0.00
y
9.99
0.01
± (3%rdg + 3dgt)
10.0
y
20.0
0.1
20.0
y
99.9
± 5%rdg
100
y
500
1
1000
0.00
y
9.99
0.01
± (3%rdg + 3dgt)
10.0
y
20.0
0.1
20.0
y
99.9
± 5%rdg
100
y
1000
1
Test voltage accuracy:
(-0%
y
+25%) of Un
Test current:
>1mA (up to Un/1mA)
Short-circuit current:
< 15mA
Discharging object on test:
Internal resistance of 2M
:
after finishing measurement
Detection of external voltage
UEXT lim = 10VAC (between RISO+ and RISO- terminals
UEXT lim = 50 VAC (between RISO+ and RISO- terminals
UEXT lim = 50VAC approx. (between any RISO terminal and GND)
UEXT lim = -10VDC (between RISO+ and RISO- terminals)
DIELECTRIC WITHSTANDING TEST (DIELECTRIC)
Nominal test voltage
UN [V]
Resolution
(V)
Accuracy
Overload
protection
250
y
800
10 ±3%U
N
CAT III 300 V
810
y
2500
2510
y
5100
Nominal test voltage UN:
Adjustable 250
y
5100 V, 50/60 Hz in steps of 10V
Distortion of test voltage:
Crest factor CF = 1.414 ± 5%
Measurement modes:
MANUAL, RAMP (timer), BURN or PULSE (duration of 3
measurement cycles:: 60ms@50Hz, 50ms@60Hz
Output power:
500VA (@ 5100V)
Discharging current IAPP
Range [mA]
Resolution [mA]
Accuracy
0
y
200
1
±(3%rdg + 2mA)
Discharging current IREAL
Range [mA]
Resolution [mA]
Accuracy
0
y
110
1
± (3%rdg + 4mA)
Nominal breaking current (IAPP or IREAL):
Adjustable 1
y
110 mA in steps of 1mA
Short-circuit current:
>200mA
Breaking time:
<30ms
Summary of Contents for FULLTEST3
Page 1: ... Copyright HT ITALIA 2020 Release EN 2 01 13 01 2020 FULLTEST3 User manual ...
Page 2: ......