FULLTEST3
EN - 77
6.13. EXECUTION OF AN AUTOTEST
The instrument allows you to execute Autotest sequences predefined in the main
Menu (see § 5.7).
1. Press the
FUNC
key and select the
AUTO TEST
function. The screen of is shown
on the display
Fig. 76 : AUTO TEST Function - Initial screen
2. Use the
S
or
T
arrow keys to select the desired Autotest (ex: Insulation
measurement with 500VDC test voltage and 3 tests included) and touch the
SELECT
key or directly the corresponding line. The "0 / X" message in which X =
number of tests included in the Autotest indicates that no internal test has been
performed. The screen of Fig. 76 - right side is shown on the display
3. Connect the instrument to the first circuit under test (ex: consider the Insulation
measurement of § 6.3)
4. Press the
START/STOP
key twice to activate the "
01
" test of the Autotest
5. Press the
START/STOP
key again to end the "
01
" test. The instrument will save
the first partial result of the Autotest and automatically prepares itself for the
execution of the next "
02
" test (see Fig. 77)
6. Press the
COMMENT
key to include a possible comment to the 01 test
Fig. 77 : AUTO TEST Function - Partial result test 01
7. Connect the test leads to the second circuit under test
8. Press the
START/STOP
key twice to activate the "
02
" test of the Autotest
9. Press the
START/STOP
key again to end the "
02
" test. The instrument will save
the first partial result of the Autotest and automatically prepares itself for the
execution of the next "
03
" test (see Fig. 78)
10. Press the
COMMENT
key to include a possible comment to the 02 test
Summary of Contents for FULLTEST3