FULLTEST3
EN - 23
NOTE
The EN50191 regulation also provides, when performing dielectric strength tests, the
restriction of testing area of the on access and use of lamps for the reporting of
hazardous conditions. In this regard the following optional accessories are available:
¾
FT3SFTSW
Æ
safety contact (with connector and cable) to fix at the door of
testing area zone
¾
FT3R-GLP
Æ
Red/Green Warning Lamp for FULLTEST3 with Hardware 70 or
higher (from S/N:
16101107
)
¾
FT3REDLP
Æ
Red Warning Lamp for FULLTEST3 with Hardware version lower
Hardware 70
In order to set the status of safety input in DIELECTRIC function the SAFETY INPUT
menu shall be used. The safety input may be enabled or disabled. Safety input
disabled: DIELECTRIC test is active regardless of the safety input condition (safety
switch may be closed or opened or not connected at all). Safety input enabled:
DIELECTRIC test is active only if safety input condition is sufficient (safety switch
must be closed). This safety input status does not influence to any other function
except DIELECTRIC.
5.6. SOUND MENU
In order to turn the acoustic signals off/on SOUND menu shall be used.
OFF
SOUND
ON
Fig. 17 : SOUND Menu