FULLTEST3
EN - 31
Item Description
11
Real time clock (hh.mm.ss).
12
Unit of measurement result (
:
)
13
Set measurement time (in TIMER mode only)
14
Measurement result status (symbol
in green color - result OK,
symbol
in red color - result not OK or symbol
in yellow color –
result OK, but measurement current too low
11. Test result will currently be displayed in green (result lower than or equal to set
limit value) or in red colour (result higher than set limit value). Final result will be
equipped with green
symbol and with beep-beep sound (result OK) or with red
symbol and with longer beep sound (result not OK) or with yellow
symbol
and with beep-beep sound (result OK, but measurement current was too low)
12. Save the test result by pressing the
SAVE
key (see § 7.1)
CAUTION
x
Max external voltage between two R
PE
or between two SENSE
test terminals is 10 VAC, no DC external voltage is allowed! In
case of higher external voltage fuse F4 (
T20A/500V, 6.3×32 mm
)
may blow
x
Measurement time in MANUAL mode is limited to 5min
x
The measurement time can be set from 2s÷60min
independently by the selected test current except for 25A
measurement in which the measurement time is from 00:02 to
05:00 (2s to 5min)