FULLTEST3
EN - 76
Meaning of symbols on the display
Item Description
1
Selected function
2
Sub-result, mains voltage U
L/N
3
Measurement screen touch-screen key
4
MODE
touch-screen key to select measurement mode (SOCKET)
Currently selected mode is displayed on the bottom of the key
5
LIMIT
touch-screen key to select limit leakage current. Currently selected
value is displayed on the bottom of the key
6
L POS
touch-screen key to select the position of phase terminal on
schuko socket during the measurement. Currently selected position is
displayed on the bottom of the key
7
Measurement result (in green colour - result OK, in red colour - result not
OK).
8
Real time clock (hh.mm.ss)
9
Unit of the test result
10
Measurement result status (symbol
in green colour - result OK,
symbol
in red colour - result not OK).
9. Save the test result by pressing the
SAVE
key (see § 7.1)
CAUTION
x
Measurement at both phase positions (phase at left terminal and
phase at right terminal) must be carried out and the highest
value must be evaluated
x
Switch on the UUT in order total leakage current to be measured
x
In case of overloaded test socket the fuse F1 or F2 (both
T16A/250V) may blow
x
Do not use test schuko socket except for measurement purpose
x
Measurement time (SOCKET) is limited to 60 min
6.12.1. Anomalous
situations
The following specific information can be shown on the display during measurement
Information displayed
Description
I
PE
CURRENT OVERRANGE!
If I
PE
current is higher than 10A for 10s, the
measurement will be stopped and this message
will appear