FULLTEST3
EN - 60
Item Description
6
I
'
N
touch-screen key to select nominal differential current. Currently
selected value is displayed in the bottom of the key
7
Measurement result (in green color - result OK, in red color - result
not OK).
8
Real time clock (hh.mm.ss)
9
Unit of the test result
10
Measurement result status (symbol
in green color - result OK,
symbol
in red color - result not OK)
7. Test result will be displayed after elapsing measurement time in green colour
equipped with green
symbol and with beep-beep sound if it is lower than or
equal to limit value (see the explanation of limit value above). If the result is higher
than limit value, then it will be displayed in red color equipped with red
symbol
and with longer acoustic sound
8. Save the test result by pressing the
SAVE
key (see § )
CAUTION
In case voltage U
L/PE
within required range 100
y
265 V is present
between L and PE test terminals (also displayed) but there is no
READY message displayed, check if energizing socket is correctly
earthed
6.7.2. Anomalous situations
The following specific information can be shown on the display during measurement
Information displayed
Description
VOLTAGE OUT OF RANGE!
Input voltage U
L/PE
out of required range
100
y
265 V after pressing the START key.
CONTACT VOLTAGE > 50V
or
CONTACT VOLTAGE > 25V
Contact voltage higher than selected limit
value, probably due to too high loop
resistance.
MEASUREMENT FAILED!
Measurement current was interrupted either
due to disconnection of test leads or due to
increase of loop resistance.
FUSE F3!
Fuse F3 is blown.
HOT!
Internal circuitry is overheated.
Wait to cool it down!