FULLTEST3
EN - 88
TECHNICAL SPECIFICATIONS
10.5. TECHNICAL CHARACTERISTICS
Accuracy calculated as ±[%r (num.dgt * resolution)] at 23°C ± 5°C, <60%RH
CONTINUITY OF PE CONDUCTOR (RPE-2WIRE, 200mA)
Range [
:
]
Resolution [
:
]
Accuracy
Overload
protection
0.00
y
19.99
0.01
± (3%rdg + 3dgt)
CAT III 300 V
20.0
y
200.0
0.1
Open-circuit test voltage:
approx. 4.5VAC
Short-circuit test current:
<0.6A (standard test leads)
Test current:
>200mA (standard test leads and external resistance < 20
:
)
Range of test current:
10mA
y
255mA
Accuracy of test current:
±(3%rdg. + 2dgt)
Limit value:
0.01
:
y
200.0
:
adjustable
Timer on measurement:
2s ÷ 60min programmable
Measurement principle:
2-Wire connection
Test lead calibration:
up to 5.00
:
Protection: Fuse
F4
Detection of external voltage
UEXT lim = 3VAC (between two RPE or between two SENSE terminals
UEXT lim = 10VAC (between two RPE or between two SENSE terminals
UEXT lim = 30VAC approx. (between any RPE/SENSE terminal and GND
CONTINUITY OF PE CONDUCTOR (RPE-2WIRE, 25A)
Range [
:
]
Resolution [
:
]
Accuracy
Overload
protection
0.000
y
1.999
0.001
± (3%rdg. + 3dgt)
CAT III 300 V
2.00
y
20.00
0.01
Open-circuit test voltage:
approx. 4.5VAC
Short-circuit test current:
<30A (standard test leads)
Test current (25A range):
> 25A (
standard test leads and external resistance < 0.1
:
)
>10A
(
standard test leads and external resistance < 0.5
:
)
Range of test current:
0.2A
y
30.0A
Accuracy of test current:
± (3%rdg. + 1dgt)
Limit value:
0.01
:
y
20.00
:
adjustable OR calculation through loop impedance OR
calculation through wire length
Timer on measurement:
2s ÷ 60min programmable (test current > 10A)
2s ÷ 5min programmable (test current > 25A)
Measurement principle:
Two-wire connection
Test lead calibration:
Up to 5.00
:
Protection: Fuse
F4
Detection of external voltage
Yes (see RPE-2WIRE, 200mA)
CONTINUITY OF PE CONDUCTOR (RPE-4WIRE, 25A)
Range [
:
]
Resolution [
:
]
Accuracy
Overload
protection
0.000
y
1.999
0.001
± (3%rdg. + 3dgt)
CAT III 300 V
2.00
y
20.00
0.01
Open-circuit test voltage:
approx. 4.5VAC
Short-circuit test current:
<30A (standard test leads)
Test current (25A range):
>25A (
standard test leads and external resistance < 0.1
:
)
>10A
(
standard test leads and external resistance < 0.5
:
)
Range of test current:
0.2A
y
30.0A
Accuracy of test current:
± (3%rdg. + 1dgt)
Limit value:
(see RPE-2WIRE 25A)
Timer on measurement:
2s ÷ 5min programmable
Measurement principle:
4-Wire connection
Protection: Fuse
F4
automatically
detected
Detection of external voltage
Yes (see RPE-2WIRE, 200mA)