FULLTEST3
EN - 72
7. Start the measurement by pressing the
START/STOP
key. The measurement will
start to run and will be stopped after pressing the
START/STOP
key again.
Measurement result will be currently displayed in green color if it is lower than or
equal to set limit value or in red color if it is higher than set limit value. Final result
will be equipped with green
symbol and with acoustic sound if it is OK or with
red
symbol and with longer acoustic sound if it is not OK. See the display
outlook with test result in Fig. 70
Fig. 70 : Visualization of ICLAMP test result
Meaning of symbols on the display
Item Description
1
Selected function
2
Measurement screen touch-screen key
3
RANGE
touch-screen key to select measurement range (0
y
1000
mA, 0
y
100 A or 0
y
1000 A
4
LIMIT
touch-screen key to select limit current value inside each
measurement range
5
Measurement result (in green color - result OK, in red color - result
not OK)
6
Real time clock (hh.mm.ss).
7
Measurement unit of result
5
Measurement result status (symbol
in green color - result OK,
symbol
in red color - result not OK).)
8. Save the test result by pressing the
SAVE
key (see § 7.1)
CAUTION
x
Max input voltage is 10 V, one terminal is grounded
x
Measurement time is limited to 60 min
6.11.1. Anomalous
situations
The following specific information can be shown on the display during measurement
Information displayed
Description
CURRENT OVERRANGE
If IL current is higher than 10 A for 10s, the
measurement will be stopped and this message will
appear
Summary of Contents for FULLTEST3