FULLTEST3
EN - 22
5.5.2. EN50191 Menu
CAUTION
The “EN50191” menu is available only for instrument with FW
version B30.Mxx.Vxx or higher
In order to select the status of safety level in DIELECTRIC function the
EN50191
menu shall be used (it replace “old” SAFETY” button). The operation according to
EN50191 guideline may be enabled or disabled.
This safety level status does not
influence any other function except DIELECTRIC
DISCLAIMER
x
Since the instrument could be used for applications where
EN50191 isn’t required, the instrument is delivered with this
option DISABLED
x
The user is warned to enable this option if his test procedure
require additional safety precautions (typically when test
current is
3mA)
x
The producer declines any responsibility if the EN50191 option
isn’t correctly set according to safety procedures
DISABLED
ENABLED
SAFETY INPUT
Fig. 16 : SAFETY INPUT Menu
Operation according to EN50191 standard is enabled:
DIELECTRIC function is
accessible only if the operator enter the (not modified)
8314
password and (for test
voltage
100V) the “SAFETY INPUT” contact is closed. The password is required
only during the execution of the first dielectric test after switch on or the enabled of
EN50191 parameter.
Operation according to EN50191 standard is disabled:
DIELECTRIC function is
accessible regardless of safety input condition (safety switch may be closed or open
or not connected at all) and no entry password is required when starting the
DIELECTRIC test.