FULLTEST3
EN - 68
Item Description
7
L POS
touch-screen key to select the position of phase terminal on
schuko socket during the measurement. Currently selected position
is displayed on the bottom of the key
8
Measurement result (in green colour - result OK, in red colour - result
not OK).
9
Real time clock (hh.mm.ss).
10
Unit of the test result
11
Set measurement time
12
Measurement result status (symbol
in green colour - result OK,
symbol
in red colour - result not OK)
10. Save the test result by pressing the
SAVE
key (see § 7.1)
CAUTION
x
Measurement at both phase positions (phase at left terminal and
phase at right terminal) must be carried out when leakage
current IPE is measured and the highest value must be
evaluated
x
Switch on the UUT in order full power of the unit and total
leakage current to be measured
x
In case of overloaded test socket the fuse F1 or F2 (both
T16A/250V) may blow
x
Do not use test schuko socket except for measurement purpose
6.9.1. Anomalous situations
The following specific information can be shown on the display during measurement
Information displayed
Description
I
PE
> 3.5 mA
Leakage current I
PE
is higher than 3.5 mA what
may be dangerous for the operator. The
message will appear always when the current
exceeds the 3.5mA threshold and will
automatically disappear after 10 s. For more
obvious warning the message is accompanied
with audio beep-beep signal.
I
PE
CURRENT OVERRANGE!
If I
PE
current is higher than 10 A for 10 s, the
measurement will be stopped and this message
will appear.
I
L
CURRENT OVERRANGE!
If I
L
current is higher than 16 A for 10 s, the
measurement will be stopped and this message
will appear.