9. UPDATING FIRMWARE OF THE INSTRUMENT
It is possible to update the internal firmware (FW) of the instrument by using a USB
pen drive. Proceed as follows:
1.
Update the
TopView
management software supplied at the latest available
version
2.
Insert a USB drive with a
maximum size of 64GB FAT32 formatted
into the PC
3.
Download the latest version of the FW from the "PC
Æ
instrument connection"
section of TopView
4.
Run the downloaded file "
90550_PENDRIVE_FW_UPG_setup.exe
" which will
load the latest version of the FW inside the USB pen drive
5.
Remove the pen drive and connect it to the instrument's
USB2
or
USB3
inputs
6.
Confirm the FW update request message shown on the display
7.
Wait for the instrument display to return to the home screen and remove the USB
pen drive. The new FW has been installed
8. Check the installed version (see § 5.4)