Parameter Description
Value
Utest NOM
Nominal test voltage
250V ÷ 5100VAC
MODE Measurement
mode
Manual, Burn, Pulse
Ramp 75%, Ramp 50%
TIMER
Measurement mode (Ramp only)
00:01 ÷ 10:00 (1s ÷ 10min)
LIMIT
Discharge current limit threshold
1mA ÷ 110mA
CHAR
Typology of discharge current
I
APP
or I
REAL
(see §)
Table 4 : Setup parameters of DIELECT function
6.4.1. Measurement modes
The instrument allows the selection of the following operating modes:
¾
Manual
mode
Æ
The test voltage is maintained constantly until the
START/STOP
key is pressed (see Fig. 36). The measured discharge current is compared with
the set limit value and the result is stored in memory
¾
Burn
mode
Æ
The test voltage is maintained constantly until the
START/STOP
key is pressed (see Fig. 36) but the result is NOT compared with any limit and is
NOT stored in the memory (functional test)
Fig. 36 : DIELECTRIC test in Manual or Burn mode
¾
75% Ramp
mode (
single ramp
)
Æ
When the
START/STOP
key is pressed, the
test voltage rises up to 75% of the nominal voltage then it takes 5s to reach the
nominal value. It is subsequently maintained for a time defined by a
programmable timer (see Fig. 37)
Fig. 37 : DIELECTRIC test in 75% Ramp mode