4. To read data directly from the CVT hardware table in an HP VEE
program, select an HP VEE direct I/O panel for the HP E1313/E1413.
In this panel select:
READ
MEMORY CVT
X
ACCESS MODE
Increment On
ARRAY 1D SIZE 64
This sets up HP VEE to read the CVT Current Value Memory that in
step 2 we named “ CVT” and store the results in variable X.
The Access Mode “ Increment on” means that successive memory
locations will be read to satisfy the number of readings that are
specified. In this case, 64 readings have been specified which is the
entire CVT table. This results in channel zero as the first element in
the 1D array and channel 63 as the 64th element.
390 Using HP VEE with the HP E1313/E1413
Appendix E